*** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 01 Mar 2004 17:44:58 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Mo, 01 Mar 2004 17:45:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040301 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.56 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040301 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1757.7 1757.9 827.6 0.12 0.12 2.12 2 6.16 3510.4 3510.0 1646.9 -0.12 -0.14 2.13 3 12.31 7024.1 7023.9 3311.9 -0.13 -0.14 2.12 4 18.47 10539.5 10539.8 5103.2 -0.12 -0.12 2.07 5 27.70 15816.0 15815.8 7674.7 -0.09 -0.09 2.06 6 36.93 21099.2 21100.6 10162.4 -0.04 -0.04 2.08 7 46.16 26385.3 26387.8 12713.2 -0.00 0.01 2.08 8 61.54 35219.8 35221.9 16603.9 0.11 0.11 2.12 9 76.93 44021.3 44023.5 20795.6 0.10 0.10 2.12 10 92.31 52871.3 52875.2 24518.5 0.18 0.19 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3767 to 113306 e- Average count rate measured: 572 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay