*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 22 Feb 2004 16:46:20 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 22 Feb 2004 16:46:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040222 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.74 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040222 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1753.8 1753.4 825.7 0.10 0.11 2.12 2 6.16 3502.7 3501.4 2065.6 -0.13 -0.15 1.70 3 12.31 7007.6 7008.8 3309.8 -0.14 -0.12 2.12 4 18.47 10514.5 10514.1 4976.4 -0.12 -0.13 2.11 5 27.70 15778.7 15779.3 7631.9 -0.09 -0.09 2.07 6 36.93 21051.8 21052.2 10129.3 -0.03 -0.03 2.08 7 46.16 26324.9 26326.8 12578.1 0.01 0.01 2.09 8 61.54 35135.9 35139.0 16742.6 0.11 0.11 2.10 9 76.93 43921.8 43922.4 20940.1 0.11 0.10 2.10 10 92.31 52739.8 52747.8 24440.0 0.17 0.18 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3760 to 113096 e- Average count rate measured: 570 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.24 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay