*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 08 Feb 2004 14:23:58 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 08 Feb 2004 14:24:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040208 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.35 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1753.4 1753.7 830.1 0.13 0.13 2.11 2 6.16 3501.2 3501.9 1668.0 -0.14 -0.13 2.10 3 12.31 7007.1 7007.4 3329.9 -0.13 -0.13 2.10 4 18.47 10513.1 10513.4 5029.8 -0.13 -0.13 2.09 5 27.70 15777.3 15777.6 7625.0 -0.09 -0.09 2.07 6 36.93 21049.2 21049.2 10126.8 -0.04 -0.04 2.08 7 46.16 26324.0 26322.8 12776.9 0.01 -0.00 2.06 8 61.54 35134.4 35135.1 16770.5 0.11 0.11 2.10 9 76.93 43917.8 43923.2 20579.2 0.10 0.11 2.13 10 92.31 52741.7 52743.6 24534.8 0.18 0.18 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3753 to 112869 e- Average count rate measured: 570 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.19 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay