*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 01 Feb 2004 18:52:32 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 01 Feb 2004 18:53:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040201 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.87 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040201 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1679.7 1680.0 793.3 0.11 0.11 2.12 2 6.16 3356.7 3357.0 1590.2 -0.10 -0.10 2.11 3 12.31 6714.6 6715.0 3317.0 -0.15 -0.15 2.02 4 18.47 10078.6 10078.3 4881.7 -0.10 -0.11 2.06 5 27.70 15119.5 15120.8 7261.6 -0.10 -0.10 2.08 6 36.93 20173.0 20175.1 9660.7 -0.04 -0.04 2.09 7 46.16 25222.8 25224.3 12150.1 -0.02 -0.02 2.08 8 61.54 33680.8 33686.2 16140.0 0.13 0.13 2.09 9 76.93 42077.4 42083.3 20020.7 0.07 0.07 2.10 10 92.31 50568.9 50575.5 23593.0 0.22 0.22 2.14 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3574 to 107599 e- Average count rate measured: 547 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.22 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.011 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay