*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 23 Jan 2004 13:20:17 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 23 Jan 2004 13:21:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040123 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.61 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040123 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1771.8 1772.5 837.5 0.11 0.12 2.12 2 6.16 3541.3 3540.9 1660.8 -0.10 -0.11 2.13 3 12.31 7083.3 7083.4 3353.0 -0.15 -0.15 2.11 4 18.47 10630.8 10631.0 5098.4 -0.11 -0.12 2.09 5 27.70 15951.5 15952.7 7776.1 -0.09 -0.09 2.05 6 36.93 21283.3 21284.3 10184.1 -0.03 -0.03 2.09 7 46.16 26609.8 26612.8 12679.2 -0.01 -0.01 2.10 8 61.54 35528.3 35528.0 17128.5 0.12 0.12 2.07 9 76.93 44383.6 44390.2 20918.4 0.06 0.07 2.12 10 92.31 53329.7 53334.2 24452.1 0.19 0.19 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3813 to 114744 e- Average count rate measured: 577 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.02 e-/ADU RON = 4.26 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.011 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay