*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 16 Jan 2004 13:31:14 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 16 Jan 2004 13:32:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040116 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 356.40 stddev = 1.96 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040116 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1747.0 1747.0 820.8 0.12 0.14 2.13 2 6.16 3487.3 3486.7 1652.8 -0.16 -0.17 2.11 3 12.31 6981.0 6980.4 3338.7 -0.12 -0.12 2.09 4 18.47 10471.6 10471.4 4995.8 -0.13 -0.13 2.10 5 27.69 15717.2 15716.6 7625.4 -0.08 -0.08 2.06 6 36.93 20966.0 20965.3 10153.1 -0.04 -0.04 2.06 7 46.16 26221.6 26223.1 12487.2 0.01 0.02 2.10 8 61.54 34992.7 34993.3 16747.8 0.10 0.10 2.09 9 76.93 43754.5 43756.2 20619.1 0.13 0.13 2.12 10 92.31 52521.2 52525.1 24638.4 0.16 0.16 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3714 to 111673 e- Average count rate measured: 568 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.13 +/- 0.01 e-/ADU RON = 4.18 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay