*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 09 Jan 2004 16:38:52 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 09 Jan 2004 16:39:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040109 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 355.03 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040109 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1686.9 1686.9 804.4 0.12 0.12 2.10 2 6.16 3368.1 3368.7 1583.0 -0.14 -0.13 2.13 3 12.31 6740.8 6741.1 3238.1 -0.13 -0.13 2.08 4 18.47 10114.1 10114.0 4800.6 -0.12 -0.13 2.11 5 27.70 15176.7 15177.2 7353.1 -0.09 -0.09 2.06 6 36.93 20249.0 20249.2 9760.6 -0.03 -0.04 2.07 7 46.16 25321.6 25321.4 12299.9 0.00 -0.00 2.06 8 61.54 33797.8 33797.5 16201.7 0.11 0.10 2.09 9 76.93 42245.2 42249.8 20357.3 0.10 0.11 2.08 10 92.31 50737.9 50742.0 23874.8 0.18 0.19 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3547 to 106703 e- Average count rate measured: 549 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.10 +/- 0.01 e-/ADU RON = 4.17 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay