*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 02 Jan 2004 16:29:59 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 02 Jan 2004 16:30:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040102 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.98 stddev = 1.98 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_46_040102 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 3.08 1738.4 1738.1 825.9 0.13 0.13 2.10 2 6.16 3470.3 3471.3 1594.4 -0.14 -0.12 2.18 3 12.31 6944.6 6945.0 3327.9 -0.14 -0.14 2.09 4 18.47 10420.6 10419.7 5031.4 -0.12 -0.14 2.07 5 27.70 15637.7 15638.6 7569.6 -0.09 -0.09 2.07 6 36.93 20863.6 20862.0 10104.3 -0.03 -0.04 2.06 7 46.16 26088.8 26089.8 12569.7 0.00 0.00 2.08 8 61.54 34821.4 34823.9 16547.0 0.10 0.11 2.10 9 76.93 43529.0 43531.7 20586.3 0.11 0.11 2.11 10 92.31 52275.8 52277.4 24197.6 0.18 0.18 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3723 to 111957 e- Average count rate measured: 565 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.02 e-/ADU RON = 4.23 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay