*** MIDAS logfile (05SEPpl1.2) *** We, 30 Jan 2008 15:34:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 30 Jan 2008 15:34:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 080130 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.40 stddev = 1.85 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_080130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1618.0 1617.5 771.0 0.13 0.11 2.10 2 1.09 3217.5 3217.2 1519.8 -0.26 -0.25 2.12 3 2.17 6459.7 6457.3 3047.8 0.07 0.05 2.12 4 3.25 9669.3 9669.6 4612.8 -0.11 -0.10 2.10 5 4.88 14522.7 14518.1 7305.3 0.02 0.01 1.99 6 6.50 19344.7 19347.3 9603.0 -0.09 -0.07 2.01 7 8.13 24212.1 24208.1 12223.1 0.04 0.04 1.98 8 10.84 32265.5 32266.1 16074.6 0.00 0.01 2.01 9 13.55 40364.8 40362.9 20241.7 0.08 0.08 1.99 10 16.25 48447.7 48447.2 24081.2 0.10 0.11 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3225 to 96589 e- Average count rate measured: 2978 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.01 e-/ADU RON = 3.70 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.25 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay