*** MIDAS logfile (05SEPpl1.2) *** Sa, 01 Dec 2007 12:59:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 01 Dec 2007 13:00:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 071201 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 299.19 stddev = 1.87 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_071201 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1596.5 1599.5 766.8 0.12 0.07 2.08 2 1.09 3182.5 3184.9 1547.2 -0.21 -0.24 2.06 3 2.17 6389.9 6389.8 3069.0 0.08 0.05 2.08 4 3.25 9566.3 9568.5 4726.5 -0.11 -0.10 2.02 5 4.88 14368.9 14369.0 7132.4 0.02 0.02 2.01 6 6.50 19144.7 19145.7 9417.8 -0.08 -0.06 2.03 7 8.13 23952.7 23957.7 11858.9 0.02 0.05 2.02 8 10.84 31930.4 31929.9 15826.5 -0.00 0.01 2.02 9 13.55 39942.0 39944.2 19362.7 0.07 0.09 2.06 10 16.25 47946.0 47942.9 23707.0 0.10 0.11 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3247 to 97417 e- Average count rate measured: 2947 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.03 +/- 0.01 e-/ADU RON = 3.81 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay