*** MIDAS logfile (05SEPpl1.2) *** Su, 26 Nov 2006 12:27:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 26 Nov 2006 12:27:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061126 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.65 stddev = 1.88 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_061126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1715.9 1716.0 832.1 0.07 0.05 2.06 2 1.09 3419.3 3421.7 1633.9 -0.21 -0.19 2.09 3 2.17 6859.9 6863.1 3329.2 0.05 0.08 2.06 4 3.25 10272.8 10272.5 5076.9 -0.10 -0.11 2.02 5 4.88 15424.7 15427.9 7599.8 0.01 0.02 2.03 6 6.50 20558.3 20553.5 10282.2 -0.05 -0.07 2.00 7 8.13 25720.7 25721.4 12708.6 0.04 0.04 2.02 8 10.84 34282.5 34281.7 17278.0 0.01 0.01 1.98 9 13.55 42877.7 42877.4 21414.5 0.07 0.06 2.00 10 16.25 51474.3 51475.9 25828.8 0.11 0.11 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3411 to 102330 e- Average count rate measured: 3164 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.01 e-/ADU RON = 3.73 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay