*** MIDAS logfile (05SEPpl1.2) *** Fr, 10 Nov 2006 15:35:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 10 Nov 2006 15:35:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061110 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.42 stddev = 1.88 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_061110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1700.9 1703.4 824.5 0.09 0.18 2.06 2 1.09 3392.3 3389.9 1640.5 -0.21 -0.25 2.07 3 2.17 6810.9 6805.4 3327.3 0.10 0.03 2.05 4 3.25 10194.7 10194.4 5025.6 -0.10 -0.11 2.03 5 4.88 15314.3 15310.9 7728.4 0.04 0.01 1.98 6 6.50 20400.6 20403.5 10080.0 -0.07 -0.06 2.02 7 8.13 25527.9 25528.3 12719.3 0.03 0.03 2.01 8 10.84 34024.2 34024.8 16734.9 -0.00 0.00 2.03 9 13.55 42550.9 42560.8 21056.8 0.04 0.07 2.02 10 16.25 51086.0 51085.4 25273.8 0.09 0.09 2.02 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3440 to 103251 e- Average count rate measured: 3141 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.01 e-/ADU RON = 3.80 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay