*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 12:29:16 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 12:29:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061018 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.77 stddev = 1.91 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1702.0 1705.7 822.0 0.02 0.15 2.07 2 1.09 3395.2 3396.4 1632.1 -0.20 -0.19 2.08 3 2.17 6818.4 6814.4 3353.5 0.09 0.03 2.03 4 3.25 10204.4 10207.9 5085.6 -0.11 -0.10 2.01 5 4.88 15327.6 15326.0 7753.3 0.01 -0.01 1.98 6 6.50 20424.9 20425.9 10248.0 -0.06 -0.08 1.99 7 8.13 25558.0 25559.3 12864.0 0.04 0.03 1.99 8 10.84 34065.1 34066.1 17188.2 0.01 -0.00 1.98 9 13.55 42606.8 42609.8 21551.6 0.07 0.06 1.98 10 16.25 51152.4 51151.2 25691.8 0.12 0.10 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3373 to 101263 e- Average count rate measured: 3144 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.00 e-/ADU RON = 3.79 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay