*** MIDAS logfile (05SEPpl1.2) *** Fr, 13 Oct 2006 16:44:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 13 Oct 2006 16:45:00 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 061013 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.11 stddev = 1.85 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_061013 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1717.5 1718.8 824.0 0.14 0.04 2.08 2 1.09 3423.7 3423.6 1639.1 -0.15 -0.21 2.09 3 2.17 6867.6 6868.3 3375.6 0.07 0.07 2.03 4 3.25 10278.2 10282.2 5099.1 -0.12 -0.10 2.02 5 4.88 15440.1 15440.0 7599.1 0.00 0.01 2.03 6 6.50 20573.1 20573.6 10278.2 -0.08 -0.06 2.00 7 8.13 25741.3 25742.4 12915.3 0.02 0.04 1.99 8 10.84 34313.4 34316.0 17474.7 -0.00 0.02 1.96 9 13.55 42908.9 42915.5 21248.7 0.03 0.07 2.02 10 16.25 51521.6 51523.5 25697.0 0.09 0.12 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3431 to 102898 e- Average count rate measured: 3167 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.01 e-/ADU RON = 3.69 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay