*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 13:57:10 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 13:57:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060906 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.02 stddev = 1.87 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1721.4 1721.1 852.5 0.05 0.03 2.02 2 1.09 3432.0 3432.8 1668.0 -0.19 -0.17 2.06 3 2.17 6884.2 6881.0 3359.4 0.08 0.05 2.05 4 3.25 10304.5 10303.6 5101.0 -0.11 -0.12 2.02 5 4.88 15477.2 15476.4 7690.1 0.03 0.03 2.01 6 6.50 20620.5 20622.5 10275.1 -0.06 -0.05 2.01 7 8.13 25800.6 25799.6 12618.1 0.04 0.04 2.04 8 10.84 34389.2 34387.4 16922.4 0.01 0.01 2.03 9 13.55 43009.3 43011.1 22188.2 0.06 0.07 1.94 10 16.25 51630.4 51634.3 25978.2 0.10 0.12 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3396 to 101881 e- Average count rate measured: 3174 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.97 +/- 0.02 e-/ADU RON = 3.69 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay