*** MIDAS logfile (05SEPpl1.2) *** Th, 31 Aug 2006 13:35:04 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 31 Aug 2006 13:35:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060831 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.19 stddev = 1.90 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060831 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1721.7 1718.9 819.8 0.11 0.03 2.10 2 1.09 3428.2 3429.3 1615.4 -0.21 -0.16 2.12 3 2.17 6875.1 6874.1 3242.0 0.06 0.05 2.12 4 3.25 10294.6 10295.8 4972.2 -0.11 -0.09 2.07 5 4.88 15461.7 15458.6 7512.2 0.02 0.02 2.06 6 6.50 20599.9 20598.2 10058.7 -0.07 -0.07 2.05 7 8.13 25772.8 25773.9 12754.7 0.03 0.04 2.02 8 10.84 34356.1 34353.2 17115.0 0.01 0.01 2.01 9 13.55 42966.3 42964.8 20962.9 0.06 0.06 2.05 10 16.25 51587.1 51584.8 25714.6 0.11 0.12 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3457 to 103665 e- Average count rate measured: 3171 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 3.81 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay