*** MIDAS logfile (05SEPpl1.2) *** Sa, 12 Aug 2006 13:12:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 12 Aug 2006 13:12:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060812 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.44 stddev = 2.46 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060812 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.55 1715.3 1713.6 818.4 0.06 0.04 2.10 2 1.09 3420.8 3420.9 1678.3 -0.17 -0.16 2.04 3 2.17 6859.8 6859.4 3379.5 0.05 0.05 2.03 4 3.25 10273.9 10276.9 5130.5 -0.10 -0.08 2.00 5 4.88 15428.4 15428.9 7577.1 0.01 0.01 2.04 6 6.50 20557.1 20561.6 10254.8 -0.08 -0.06 2.00 7 8.13 25723.5 25724.2 12892.1 0.03 0.02 2.00 8 10.84 34297.8 34297.2 16911.0 0.03 0.02 2.03 9 13.55 42882.7 42883.1 21406.7 0.05 0.05 2.00 10 16.25 51494.8 51496.9 26499.1 0.12 0.12 1.94 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3373 to 101304 e- Average count rate measured: 3165 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.97 +/- 0.02 e-/ADU RON = 4.85 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.004 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay