*** MIDAS logfile (05SEPpl1.2) *** Fr, 11 Aug 2006 16:20:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 11 Aug 2006 16:20:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060811 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.46 stddev = 2.45 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060811 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.64 2033.4 2032.0 1001.1 0.07 0.07 2.03 2 1.28 4055.9 4052.0 1928.2 0.11 0.02 2.10 3 2.55 8064.9 8071.9 3925.7 -0.28 -0.20 2.05 4 3.83 12111.8 12111.2 5728.8 -0.10 -0.15 2.11 5 5.74 18174.3 18174.4 9146.9 -0.03 -0.07 1.99 6 7.66 24162.8 24209.9 12047.4 -0.28 -0.15 2.01 7 9.57 30297.3 30328.6 15542.6 0.04 0.08 1.95 8 12.76 40459.4 40460.9 20355.8 0.20 0.15 1.99 9 15.95 50554.9 50555.0 24943.9 0.17 0.11 2.03 10 19.14 60622.6 60686.4 30710.7 0.10 0.14 1.97 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4027 to 120176 e- Average count rate measured: 3166 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.02 e-/ADU RON = 4.86 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.24 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay