*** MIDAS logfile (05SEPpl1.2) *** We, 02 Aug 2006 16:13:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 02 Aug 2006 16:13:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060802 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.57 stddev = 2.43 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060802 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1601.2 1603.4 829.3 0.04 0.02 1.93 2 1.54 3192.9 3194.7 1535.3 -0.03 -0.04 2.08 3 3.08 6379.9 6378.7 3088.0 -0.05 -0.08 2.07 4 4.62 9567.1 9565.0 4579.7 -0.04 -0.05 2.09 5 6.93 14334.3 14334.6 7214.5 -0.13 -0.12 1.99 6 9.23 19121.3 19122.2 9745.4 -0.08 -0.06 1.96 7 11.54 23922.4 23920.2 12142.7 0.02 0.03 1.97 8 15.39 31896.6 31892.2 15696.3 0.02 0.03 2.03 9 19.23 39908.5 39907.9 20436.9 0.12 0.14 1.95 10 23.08 47889.5 47881.5 24059.2 0.13 0.13 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3165 to 94586 e- Average count rate measured: 2072 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.02 e-/ADU RON = 4.81 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay