*** MIDAS logfile (05SEPpl1.2) *** Fr, 07 Jul 2006 13:38:20 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 07 Jul 2006 13:38:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060707 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.00 stddev = 1.94 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060707 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1598.3 1598.5 750.0 0.02 0.02 2.13 2 1.54 3198.2 3197.8 1541.9 0.12 0.09 2.07 3 3.08 6380.1 6381.9 3154.3 -0.14 -0.13 2.02 4 4.62 9578.7 9581.1 4692.8 -0.01 -0.01 2.04 5 6.93 14347.0 14349.8 6987.2 -0.17 -0.16 2.05 6 9.23 19147.2 19149.6 9374.2 -0.07 -0.07 2.04 7 11.54 23936.0 23940.3 12186.1 -0.06 -0.05 1.96 8 15.39 31959.4 31960.2 16040.1 0.08 0.07 1.99 9 19.23 39906.1 39912.7 20256.2 -0.02 -0.02 1.97 10 23.08 48020.5 48030.2 23604.3 0.26 0.26 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3195 to 95997 e- Average count rate measured: 2076 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.02 e-/ADU RON = 3.89 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay