*** MIDAS logfile (05SEPpl1.2) *** Su, 21 May 2006 14:04:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 21 May 2006 14:04:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060521 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.75 stddev = 1.95 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060521 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1635.6 1635.8 768.0 0.05 0.05 2.13 2 1.54 3265.6 3264.1 1564.1 -0.00 -0.02 2.09 3 3.08 6524.5 6522.6 3158.1 -0.06 -0.07 2.07 4 4.62 9783.7 9782.6 4654.8 -0.04 -0.04 2.10 5 6.93 14658.7 14660.7 7244.1 -0.15 -0.13 2.02 6 9.23 19558.0 19559.5 9777.4 -0.07 -0.07 2.00 7 11.54 24465.6 24464.1 12544.7 0.01 0.01 1.95 8 15.39 32625.2 32627.0 16139.6 0.02 0.03 2.02 9 19.23 40808.4 40808.6 20228.0 0.09 0.10 2.02 10 23.08 48995.3 48994.2 24408.8 0.15 0.15 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3273 to 98034 e- Average count rate measured: 2120 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.01 e-/ADU RON = 3.89 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay