*** MIDAS logfile (05SEPpl1.2) *** Fr, 14 Apr 2006 15:55:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 14 Apr 2006 15:56:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060414 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 296.03 stddev = 2.02 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060414 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1661.7 1662.1 791.2 0.09 0.09 2.10 2 1.54 3309.0 3311.4 1557.6 -0.10 -0.06 2.12 3 3.08 6617.0 6615.4 3189.5 -0.06 -0.07 2.07 4 4.62 9921.2 9919.7 4826.2 -0.05 -0.06 2.06 5 6.93 14867.9 14865.5 7366.8 -0.13 -0.14 2.02 6 9.23 19832.7 19835.1 9848.8 -0.07 -0.06 2.01 7 11.54 24811.0 24810.9 12029.0 0.02 0.02 2.06 8 15.39 33081.0 33080.5 16744.2 0.02 0.02 1.98 9 19.23 41393.8 41395.1 20762.9 0.13 0.13 1.99 10 23.08 49672.9 49664.3 24916.4 0.14 0.12 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3295 to 98483 e- Average count rate measured: 2149 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.01 e-/ADU RON = 4.00 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay