*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Mar 2006 18:31:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Mar 2006 18:31:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060311 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.58 stddev = 2.02 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060311 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1695.7 1696.7 1187.6 0.12 0.02 1.43 2 1.54 3379.2 3380.7 1621.1 -0.06 -0.08 2.08 3 3.08 6754.1 6753.9 3305.1 -0.05 -0.07 2.04 4 4.62 10127.5 10128.5 4875.9 -0.04 -0.04 2.08 5 6.93 15172.5 15174.5 7327.5 -0.15 -0.12 2.07 6 9.23 20242.9 20244.1 10082.7 -0.08 -0.06 2.01 7 11.54 25325.7 25325.4 12848.0 0.02 0.04 1.97 8 15.39 33762.0 33762.0 16734.8 0.00 0.03 2.02 9 19.23 42251.7 42249.1 21676.3 0.13 0.14 1.95 10 23.08 50699.5 50700.4 25180.4 0.12 0.15 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3379 to 101013 e- Average count rate measured: 2194 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.02 e-/ADU RON = 4.02 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay