*** MIDAS logfile (05SEPpl1.2) *** Su, 05 Feb 2006 18:39:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 05 Feb 2006 18:39:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060205 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.91 stddev = 1.94 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060205 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1701.5 1701.8 822.2 0.06 0.05 2.07 2 1.54 3393.5 3393.0 1626.2 -0.04 -0.05 2.09 3 3.08 6782.1 6783.1 3235.8 -0.08 -0.07 2.10 4 4.62 10172.8 10175.7 4886.4 -0.04 -0.02 2.08 5 6.93 15245.4 15244.1 7508.3 -0.13 -0.14 2.03 6 9.23 20343.6 20340.2 9958.6 -0.05 -0.06 2.04 7 11.54 25437.2 25441.3 12852.8 0.00 0.01 1.98 8 15.39 33930.8 33932.5 16819.2 0.04 0.04 2.02 9 19.23 42435.4 42438.4 20996.0 0.10 0.10 2.02 10 23.08 50949.3 50952.0 25347.2 0.15 0.15 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3411 to 102131 e- Average count rate measured: 2204 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.01 e-/ADU RON = 3.88 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay