*** MIDAS logfile (05SEPpl1.2) *** Fr, 03 Feb 2006 17:33:25 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 03 Feb 2006 17:33:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 060203 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.13 stddev = 2.00 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_060203 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1694.6 1695.3 812.0 0.05 0.09 2.09 2 1.54 3384.3 3381.9 1628.4 -0.01 -0.05 2.08 3 3.08 6761.5 6762.0 3211.6 -0.06 -0.05 2.11 4 4.62 10141.5 10139.3 4926.0 -0.03 -0.04 2.06 5 6.93 15192.3 15193.3 7707.8 -0.15 -0.14 1.97 6 9.23 20271.3 20270.0 10126.0 -0.07 -0.07 2.00 7 11.54 25355.9 25358.1 12575.3 0.01 0.02 2.02 8 15.39 33815.0 33812.2 16791.8 0.03 0.02 2.01 9 19.23 42303.0 42301.0 21367.6 0.11 0.11 1.98 10 23.08 50772.6 50769.9 25351.7 0.13 0.13 2.00 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3372 to 101006 e- Average count rate measured: 2197 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.99 +/- 0.01 e-/ADU RON = 3.97 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.001 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay