*** MIDAS logfile (05SEPpl1.1) *** Sa, 26 Nov 2005 15:22:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.1) *** Sa, 26 Nov 2005 15:22:49 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 051126 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 299.77 stddev = 1.95 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_051126 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1686.5 1685.8 795.5 0.05 -0.01 2.12 2 1.54 3363.3 3366.0 1575.2 -0.04 -0.00 2.14 3 3.08 6718.3 6721.0 3300.9 -0.09 -0.08 2.04 4 4.62 10078.9 10079.8 4957.0 -0.03 -0.04 2.03 5 6.93 15099.5 15100.2 7463.9 -0.14 -0.14 2.02 6 9.23 20149.4 20148.8 9859.4 -0.05 -0.06 2.04 7 11.54 25196.1 25202.5 12483.2 0.00 0.01 2.02 8 15.39 33610.7 33610.2 16727.3 0.05 0.04 2.01 9 19.23 42020.8 42039.5 21323.2 0.07 0.11 1.97 10 23.08 50478.1 50480.9 24863.3 0.18 0.17 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3375 to 101038 e- Average count rate measured: 2183 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.01 e-/ADU RON = 3.90 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay