*** MIDAS logfile (03SEPpl1.2) *** Fr, 03 Jun 2005 13:47:27 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 03 Jun 2005 13:47:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#45 TAKEN ON 050603 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 299.94 stddev = 1.94 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_45_050603 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1746.9 1748.8 871.3 0.02 0.06 2.00 2 1.54 3486.9 3486.3 1603.7 0.00 -0.02 2.17 3 3.08 6960.3 6963.2 3281.6 -0.09 -0.07 2.12 4 4.62 10440.0 10440.5 5379.6 -0.05 -0.06 1.94 5 6.93 15641.1 15642.5 7752.0 -0.16 -0.16 2.02 6 9.23 20873.1 20875.3 10538.0 -0.06 -0.07 1.98 7 11.54 26112.2 26110.3 13043.0 0.03 0.01 2.00 8 15.39 34818.2 34820.6 17112.5 0.04 0.03 2.03 9 19.23 43555.8 43560.0 21780.7 0.12 0.12 2.00 10 23.08 52286.4 52288.3 26653.0 0.16 0.15 1.96 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3456 to 103377 e- Average count rate measured: 2262 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.02 e-/ADU RON = 3.83 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.000 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay