*** MIDAS logfile (03SEPpl1.2) *** Sa, 22 Jan 2005 21:29:23 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 22 Jan 2005 21:29:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 050122 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.02 stddev = 1.87 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_050122 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1673.3 1673.0 769.4 0.05 0.04 2.17 2 1.54 3340.8 3341.0 1594.7 -0.06 -0.05 2.09 3 3.08 6681.4 6677.9 3208.4 -0.07 -0.10 2.08 4 4.62 10021.7 10022.3 4870.5 -0.04 -0.03 2.06 5 6.93 15017.3 15016.7 7521.3 -0.14 -0.14 2.00 6 9.23 20035.9 20037.0 9973.6 -0.07 -0.06 2.01 7 11.54 25066.9 25063.5 12545.9 0.03 0.02 2.00 8 15.39 33429.1 33428.5 16722.9 0.04 0.05 2.00 9 19.23 41812.3 41812.2 20888.4 0.11 0.11 2.00 10 23.08 50198.9 50196.9 24683.5 0.16 0.16 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3364 to 100924 e- Average count rate measured: 2172 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.01 +/- 0.01 e-/ADU RON = 3.76 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay