*** MIDAS logfile (03SEPpl1.2) *** Su, 26 Dec 2004 16:39:51 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Su, 26 Dec 2004 16:40:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041226 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.95 stddev = 1.88 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_041226 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1667.2 1667.0 793.2 0.06 0.06 2.10 2 1.54 3329.3 3329.5 1581.9 -0.05 -0.04 2.10 3 3.08 6654.9 6654.7 3182.4 -0.06 -0.07 2.09 4 4.62 9981.1 9982.1 4786.5 -0.06 -0.05 2.09 5 6.93 14960.5 14957.6 7205.1 -0.13 -0.14 2.08 6 9.23 19958.0 19957.1 9860.0 -0.07 -0.07 2.02 7 11.54 24968.7 24966.7 12630.9 0.02 0.02 1.98 8 15.39 33293.3 33292.8 16348.9 0.02 0.02 2.04 9 19.23 41663.0 41657.1 20753.4 0.14 0.13 2.01 10 23.08 49995.4 49994.7 25292.2 0.14 0.14 1.98 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3302 to 99036 e- Average count rate measured: 2163 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.01 e-/ADU RON = 3.72 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay