*** MIDAS logfile (03SEPpl1.2) *** Tu, 07 Dec 2004 14:43:36 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Tu, 07 Dec 2004 14:43:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041207 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.25 stddev = 1.88 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_041207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1661.7 1659.6 765.0 0.09 0.02 2.17 2 1.54 3315.2 3317.7 1558.3 -0.03 0.02 2.13 3 3.08 6628.5 6627.2 3143.3 -0.07 -0.08 2.11 4 4.62 9942.5 9945.6 4824.1 -0.05 -0.02 2.06 5 6.93 14897.3 14895.3 7253.0 -0.15 -0.16 2.05 6 9.23 19877.6 19877.4 9784.1 -0.07 -0.07 2.03 7 11.54 24862.6 24862.7 12246.5 -0.00 -0.00 2.03 8 15.39 33163.8 33167.5 17334.9 0.03 0.04 1.91 9 19.23 41476.1 41476.6 20806.8 0.09 0.09 1.99 10 23.08 49806.8 49810.2 25048.5 0.16 0.17 1.99 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3262 to 97834 e- Average count rate measured: 2155 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.96 +/- 0.02 e-/ADU RON = 3.69 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay