*** MIDAS logfile (03SEPpl1.2) *** Mo, 06 Dec 2004 17:40:05 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 06 Dec 2004 17:40:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041206 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.68 stddev = 1.91 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_041206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1657.2 1659.1 778.7 0.08 0.03 2.13 2 1.54 3313.3 3311.8 1590.1 0.03 -0.03 2.08 3 3.08 6616.3 6617.9 3160.1 -0.12 -0.12 2.09 4 4.62 9933.4 9932.2 5022.9 -0.03 -0.04 1.98 5 6.93 14878.8 14880.3 7471.1 -0.17 -0.15 1.99 6 9.23 19858.3 19860.7 9743.3 -0.07 -0.05 2.04 7 11.54 24836.3 24839.8 12409.1 -0.01 0.01 2.00 8 15.39 33136.1 33133.9 16399.7 0.04 0.05 2.02 9 19.23 41429.5 41431.5 21337.8 0.07 0.09 1.94 10 23.08 49767.3 49772.4 24755.5 0.18 0.20 2.01 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3288 to 98691 e- Average count rate measured: 2153 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.98 +/- 0.02 e-/ADU RON = 3.78 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay