*** MIDAS logfile (03SEPpl1.2) *** Sa, 20 Nov 2004 18:21:36 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 20 Nov 2004 18:21:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 041120 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 298.90 stddev = 1.97 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_041120 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1679.8 1680.0 773.1 0.04 0.05 2.17 2 1.54 3352.2 3352.0 1566.1 -0.10 -0.10 2.14 3 3.08 6707.3 6708.2 3183.5 -0.05 -0.05 2.11 4 4.62 10058.8 10058.6 4836.2 -0.05 -0.05 2.08 5 6.93 15077.6 15075.9 7229.5 -0.13 -0.13 2.09 6 9.23 20115.8 20114.2 10096.1 -0.06 -0.07 1.99 7 11.54 25167.0 25167.6 12345.1 0.04 0.04 2.04 8 15.39 33552.9 33552.3 17211.7 0.03 0.03 1.95 9 19.23 41995.6 41989.6 21085.0 0.16 0.15 1.99 10 23.08 50382.6 50375.5 24704.5 0.14 0.13 2.04 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3356 to 100641 e- Average count rate measured: 2180 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.00 +/- 0.02 e-/ADU RON = 3.93 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay