*** MIDAS logfile (03SEPpl1.2) *** Fr, 10 Sep 2004 13:10:34 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 10 Sep 2004 13:10:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040910 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 299.22 stddev = 1.91 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_040910 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1593.0 1595.1 762.3 0.03 0.10 2.09 2 1.54 3183.6 3183.3 1467.4 -0.04 -0.06 2.17 3 3.08 6368.4 6365.9 3043.1 -0.06 -0.09 2.09 4 4.62 9552.1 9552.6 4523.8 -0.05 -0.05 2.11 5 6.93 14312.2 14312.9 6917.5 -0.16 -0.16 2.07 6 9.23 19101.9 19101.3 9478.5 -0.06 -0.07 2.02 7 11.54 23895.2 23892.8 11640.5 0.02 0.00 2.05 8 15.39 31868.5 31870.7 15938.3 0.04 0.04 2.00 9 19.23 39861.3 39862.6 20153.5 0.10 0.10 1.98 10 23.08 47866.7 47869.3 23198.1 0.18 0.18 2.06 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3213 to 96494 e- Average count rate measured: 2071 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.02 +/- 0.02 e-/ADU RON = 3.85 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay