*** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 25 Apr 2004 15:40:32 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Su, 25 Apr 2004 15:40:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040425 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 299.85 stddev = 1.91 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_040425 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1694.3 1694.3 795.4 0.04 0.05 2.13 2 1.54 3386.7 3386.0 1580.2 0.00 -0.03 2.14 3 3.08 6769.2 6769.6 3212.1 -0.10 -0.09 2.11 4 4.62 10156.6 10156.5 4918.1 -0.05 -0.05 2.07 5 6.93 15223.6 15223.9 7472.6 -0.14 -0.14 2.04 6 9.23 20315.1 20315.5 10086.5 -0.06 -0.06 2.01 7 11.54 25407.2 25408.2 12665.2 0.00 0.00 2.01 8 15.39 33892.1 33894.3 16789.9 0.04 0.05 2.02 9 19.23 42383.8 42387.9 20541.0 0.09 0.10 2.06 10 23.08 50907.3 50905.2 25036.8 0.18 0.18 2.03 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3444 to 103462 e- Average count rate measured: 2202 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.03 +/- 0.01 e-/ADU RON = 3.88 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay