*** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 23 Jan 2004 12:19:18 Page 1 *** MIDAS logfile (vers. 02SEPpl1.3) *** Fr, 23 Jan 2004 12:20:07 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#46 TAKEN ON 040123 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 297.92 stddev = 1.88 FIERA temperature = 273.15 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_46_040123 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 0.77 1778.1 1778.2 838.7 0.05 0.06 2.12 2 1.54 3557.4 3555.9 1677.2 0.06 0.04 2.12 3 3.08 7102.1 7102.7 3530.5 -0.12 -0.11 2.01 4 4.62 10662.6 10661.8 5132.5 -0.02 -0.03 2.08 5 6.93 15973.6 15971.4 7921.1 -0.15 -0.17 2.02 6 9.23 21315.6 21315.9 10593.7 -0.07 -0.07 2.01 7 11.54 26658.7 26660.6 13140.8 -0.01 -0.01 2.03 8 15.39 35570.0 35577.6 18039.3 0.05 0.07 1.97 9 19.23 44452.0 44459.0 22474.3 0.03 0.04 1.98 10 23.08 53420.1 53425.3 26981.7 0.18 0.18 1.98 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3501 to 105179 e- Average count rate measured: 2311 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.97 +/- 0.01 e-/ADU RON = 3.70 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.003 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay