*** MIDAS logfile (vers. 01FEBpl1.1) *** Mo, 08 Oct 2001 15:32:48 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Mo, 08 Oct 2001 15:33:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 011008 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 600.25 stddev = 3.87 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_011008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1939.3 1945.2 1790.9 0.04 0.05 1.08 2 4.40 3736.4 3739.8 3454.1 -0.02 -0.06 1.08 3 8.80 7333.0 7330.4 6635.6 -0.02 -0.10 1.11 4 13.19 10917.9 10919.0 10040.3 -0.06 -0.06 1.09 5 19.78 16301.0 16301.9 15509.6 -0.07 -0.05 1.05 6 26.38 21694.6 21698.6 19685.8 -0.06 -0.03 1.10 7 32.97 27093.2 27083.9 24908.7 -0.01 -0.02 1.09 8 43.96 36090.0 36088.7 32644.6 0.02 0.05 1.11 9 54.95 45098.0 45092.3 41581.1 0.07 0.09 1.08 10 65.94 54115.9 54106.0 49067.4 0.12 0.13 1.10 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2135 to 59482 e- Average count rate measured: 817 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.10 +/- 0.01 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.175 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay