*** MIDAS logfile (vers. 01FEBpl1.1) *** Mo, 01 Oct 2001 00:24:45 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Mo, 01 Oct 2001 00:25:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 011001 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 600.84 stddev = 3.88 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_011001 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1958.5 1960.7 1825.3 0.02 0.06 1.07 2 4.40 3758.1 3756.4 3442.0 0.03 -0.05 1.09 3 8.80 7352.4 7353.0 6664.6 -0.03 -0.03 1.10 4 13.19 10933.8 10937.4 9886.0 -0.09 -0.06 1.11 5 19.78 16322.3 16323.3 14764.6 -0.06 -0.05 1.11 6 26.38 21719.1 21717.6 19482.9 -0.04 -0.04 1.11 7 32.97 27111.9 27109.1 25488.7 -0.02 -0.02 1.06 8 43.96 36114.0 36108.0 32484.7 0.03 0.02 1.11 9 54.95 45111.8 45107.6 41502.7 0.05 0.05 1.09 10 65.94 54126.8 54129.6 49116.3 0.10 0.11 1.10 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2149 to 59367 e- Average count rate measured: 818 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.10 +/- 0.01 e-/ADU RON = 4.26 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.196 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay