*** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 23 Sep 2001 16:34:32 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 23 Sep 2001 16:35:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010923 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 599.55 stddev = 3.86 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_010923 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1951.9 1953.2 1815.1 0.00 0.04 1.08 2 4.40 3774.6 3773.0 3436.3 0.03 -0.02 1.10 3 8.80 7413.1 7412.4 6658.6 -0.05 -0.06 1.11 4 13.19 11047.1 11043.8 9964.5 -0.05 -0.07 1.11 5 19.78 16496.0 16492.5 15000.4 -0.08 -0.09 1.10 6 26.38 21962.9 21965.1 19889.7 -0.05 -0.03 1.10 7 32.97 27424.8 27425.3 25307.5 -0.03 -0.01 1.08 8 43.96 36546.7 36548.5 33227.3 0.04 0.05 1.10 9 54.95 45670.8 45665.2 41276.5 0.08 0.08 1.11 10 65.94 54796.8 54793.2 49745.5 0.11 0.12 1.10 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2152 to 60379 e- Average count rate measured: 828 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.10 +/- 0.00 e-/ADU RON = 4.25 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.157 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay