*** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 08 Sep 2001 18:10:52 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 08 Sep 2001 18:11:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010908 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 600.20 stddev = 3.86 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_010908 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1901.5 1896.4 1733.6 0.07 0.03 1.10 2 4.40 3694.8 3691.6 3363.6 -0.07 -0.05 1.10 3 8.80 7285.9 7286.9 6550.3 -0.09 -0.03 1.11 4 13.19 10867.8 10869.0 9881.7 -0.10 -0.06 1.10 5 19.78 16255.4 16248.5 14641.6 -0.04 -0.08 1.11 6 26.38 21650.6 21647.3 19533.0 -0.02 -0.03 1.11 7 32.97 27039.4 27036.9 24287.1 0.00 -0.01 1.11 8 43.96 36030.6 36033.9 32674.8 0.04 0.04 1.10 9 54.95 45028.7 45026.0 41030.1 0.08 0.06 1.10 10 65.94 54046.7 54045.5 48731.9 0.14 0.12 1.11 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2098 to 59706 e- Average count rate measured: 817 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.10 +/- 0.00 e-/ADU RON = 4.26 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.123 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay