*** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 30 Aug 2001 15:15:49 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 30 Aug 2001 15:16:34 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010830 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 599.32 stddev = 3.85 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_010830 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1942.8 1949.9 1774.2 0.02 0.11 1.10 2 4.40 3744.1 3740.7 3524.7 0.02 -0.18 1.06 3 8.80 7340.5 7341.2 6637.8 -0.06 -0.08 1.11 4 13.19 10933.4 10931.0 9817.4 -0.05 -0.07 1.11 5 19.78 16323.5 16320.7 15019.6 -0.06 -0.05 1.09 6 26.38 21724.6 21724.1 19837.7 -0.05 -0.02 1.10 7 32.97 27122.1 27120.0 24696.2 -0.03 0.00 1.10 8 43.96 36128.2 36130.3 32826.4 0.01 0.06 1.10 9 54.95 45153.0 45142.4 43120.0 0.07 0.09 1.05 10 65.94 54179.3 54168.7 50724.3 0.11 0.14 1.07 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2069 to 57578 e- Average count rate measured: 818 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.06 +/- 0.01 e-/ADU RON = 4.10 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.177 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay