*** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 23 Aug 2001 14:52:58 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 23 Aug 2001 14:53:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010823 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 599.39 stddev = 3.86 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_010823 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1948.8 1947.6 1779.9 0.06 0.07 1.09 2 4.40 3782.7 3780.8 3440.4 -0.02 -0.03 1.10 3 8.80 7444.9 7445.9 7088.5 -0.13 -0.11 1.05 4 13.19 11107.6 11108.1 9910.4 -0.09 -0.09 1.12 5 19.78 16608.5 16611.9 19370.7 -0.05 -0.03 0.86 6 26.38 22113.5 22119.9 20251.1 -0.05 -0.02 1.09 7 32.97 27618.8 27621.2 25548.6 -0.01 -0.01 1.08 8 43.96 36814.0 36806.5 33305.6 0.06 0.03 1.11 9 54.95 46007.4 46003.1 41880.0 0.09 0.08 1.10 10 65.94 55212.2 55207.8 50076.9 0.14 0.12 1.10 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2169 to 61454 e- Average count rate measured: 834 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.11 +/- 0.03 e-/ADU RON = 4.30 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.133 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay