*** MIDAS logfile (vers. 01FEBpl1.1) *** We, 08 Aug 2001 17:53:57 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** We, 08 Aug 2001 17:54:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010808 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 599.39 stddev = 3.87 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_010808 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1963.5 1956.1 1780.1 0.07 0.03 1.10 2 4.40 3808.7 3806.4 3456.1 -0.06 0.03 1.10 3 8.80 7502.1 7500.2 6779.2 -0.08 -0.06 1.11 4 13.19 11188.1 11192.9 9992.9 -0.08 -0.02 1.12 5 19.78 16722.8 16721.7 15131.4 -0.07 -0.09 1.11 6 26.38 22271.0 22273.2 20194.3 -0.05 -0.05 1.10 7 32.97 27814.1 27816.1 25104.9 -0.02 -0.03 1.11 8 43.96 37070.4 37073.4 33825.5 0.04 0.02 1.10 9 54.95 46336.6 46333.5 41627.6 0.10 0.06 1.11 10 65.94 55611.2 55611.4 50396.6 0.15 0.12 1.10 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2165 to 61448 e- Average count rate measured: 841 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.10 +/- 0.00 e-/ADU RON = 4.27 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.130 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay