*** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 14 Jul 2001 19:25:47 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 14 Jul 2001 19:26:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010714 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 600.21 stddev = 3.83 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_36_010714 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 2.20 1890.4 1894.0 1722.9 0.03 0.07 1.10 2 4.40 3685.8 3684.3 3302.6 0.02 -0.09 1.12 3 8.80 7269.9 7268.4 6553.7 -0.08 -0.12 1.11 4 13.19 10849.3 10853.9 9748.7 -0.08 -0.04 1.11 5 19.78 16226.8 16226.5 14751.4 -0.05 -0.05 1.10 6 26.38 21616.4 21610.5 19534.3 -0.02 -0.04 1.11 7 32.97 26989.7 26991.0 24657.4 -0.03 -0.01 1.09 8 43.96 35976.5 35979.6 32755.2 0.03 0.05 1.10 9 54.95 44966.5 44965.8 41137.8 0.07 0.09 1.09 10 65.94 53964.5 53966.7 49331.3 0.12 0.14 1.09 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2067 to 58941 e- Average count rate measured: 816 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.09 +/- 0.00 e-/ADU RON = 4.19 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.118 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay