*** MIDAS logfile (vers. 01FEBpl1.1) *** Tu, 30 Oct 2001 16:59:52 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Tu, 30 Oct 2001 17:00:31 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 011030 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.68 stddev = 2.49 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_011030 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1743.9 1745.4 815.4 0.13 0.15 2.14 2 8.23 3427.0 3426.6 1566.9 -0.12 -0.16 2.19 3 16.47 6809.5 6808.2 3111.3 -0.13 -0.15 2.19 4 24.70 10184.7 10186.4 4676.7 -0.16 -0.15 2.18 5 37.04 15257.0 15255.7 6917.5 -0.11 -0.12 2.21 6 49.39 20336.5 20337.2 9300.9 -0.07 -0.06 2.19 7 61.73 25417.5 25418.1 11629.8 -0.02 -0.01 2.19 8 82.31 33905.0 33899.2 15672.6 0.06 0.06 2.16 9 102.89 42417.7 42412.6 19483.7 0.18 0.17 2.18 10 123.46 50926.0 50927.1 23387.6 0.25 0.26 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3793 to 110718 e- Average count rate measured: 411 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.00 e-/ADU RON = 5.41 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.118 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay