*** MIDAS logfile (vers. 01FEBpl1.1) *** Mo, 08 Oct 2001 15:00:14 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Mo, 08 Oct 2001 15:00:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 011008 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 579.30 stddev = 2.49 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_011008 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1777.4 1774.5 822.7 0.15 0.10 2.16 2 8.23 3470.9 3471.7 1587.2 -0.15 -0.07 2.19 3 16.47 6875.9 6874.3 3139.3 -0.16 -0.17 2.19 4 24.70 10279.7 10277.3 4710.1 -0.14 -0.15 2.18 5 37.04 15385.2 15385.3 7108.8 -0.11 -0.11 2.16 6 49.39 20498.0 20500.8 9553.7 -0.08 -0.07 2.15 7 61.73 25619.6 25615.4 11954.2 -0.02 -0.03 2.14 8 82.31 34166.7 34167.7 15716.1 0.07 0.07 2.17 9 102.89 42740.5 42736.5 19588.4 0.18 0.17 2.18 10 123.46 51314.7 51308.8 23549.0 0.27 0.25 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3868 to 111744 e- Average count rate measured: 414 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.01 e-/ADU RON = 5.42 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.164 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay