*** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 23 Sep 2001 16:01:41 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 23 Sep 2001 16:02:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010923 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.67 stddev = 2.49 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010923 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1792.6 1789.8 820.0 0.15 0.12 2.19 2 8.23 3508.5 3508.0 1626.8 -0.15 -0.10 2.16 3 16.47 6957.6 6955.3 3197.1 -0.17 -0.18 2.18 4 24.70 10403.6 10403.3 4801.8 -0.17 -0.16 2.17 5 37.04 15579.3 15578.9 7121.6 -0.11 -0.11 2.19 6 49.39 20761.9 20763.4 9671.8 -0.07 -0.06 2.15 7 61.73 25949.1 25948.1 11847.3 -0.01 -0.02 2.19 8 82.31 34606.5 34607.9 16046.5 0.07 0.07 2.16 9 102.89 43298.0 43293.6 20095.0 0.19 0.18 2.15 10 123.46 51983.5 51982.3 24085.5 0.27 0.27 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3862 to 112065 e- Average count rate measured: 419 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.01 e-/ADU RON = 5.37 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.145 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay