*** MIDAS logfile (vers. 99NOV) *** We, 23 May 2001 17:00:46 Page 1 *** MIDAS logfile (vers. 99NOV) *** We, 23 May 2001 17:01:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010523 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 579.09 stddev = 2.47 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010523 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1790.5 1788.2 827.5 0.12 0.10 2.16 2 8.23 3514.9 3514.4 1590.4 -0.10 -0.06 2.21 3 16.47 6973.5 6972.9 3179.1 -0.19 -0.18 2.19 4 24.70 10436.6 10436.8 4784.7 -0.13 -0.12 2.18 5 37.04 15622.6 15624.8 7236.8 -0.14 -0.12 2.16 6 49.39 20825.6 20829.6 9687.8 -0.08 -0.06 2.15 7 61.73 26030.3 26027.4 11959.2 -0.02 -0.04 2.18 8 82.31 34731.5 34731.4 16079.2 0.10 0.09 2.16 9 102.89 43435.8 43432.2 20007.7 0.17 0.16 2.17 10 123.46 52160.3 52153.7 24141.1 0.27 0.25 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3868 to 112742 e- Average count rate measured: 421 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.16 +/- 0.00 e-/ADU RON = 5.33 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.126 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay