*** MIDAS logfile (vers. 99NOV) *** We, 09 May 2001 02:44:42 Page 1 *** MIDAS logfile (vers. 99NOV) *** We, 09 May 2001 02:45:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010509 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.65 stddev = 2.47 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010509 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1794.3 1793.6 828.3 0.12 0.10 2.17 2 8.23 3526.2 3526.5 1593.8 -0.09 -0.07 2.21 3 16.47 6999.6 7003.8 3216.1 -0.18 -0.12 2.18 4 24.70 10473.9 10473.4 4838.6 -0.16 -0.17 2.16 5 37.04 15685.3 15685.4 7213.4 -0.14 -0.14 2.17 6 49.39 20911.9 20912.9 9811.9 -0.07 -0.07 2.13 7 61.73 26140.2 26137.7 11892.0 -0.01 -0.02 2.20 8 82.31 34867.3 34863.2 16188.4 0.07 0.06 2.15 9 102.89 43625.5 43616.8 20037.7 0.20 0.17 2.18 10 123.46 52374.4 52365.3 24639.2 0.27 0.25 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3845 to 112249 e- Average count rate measured: 423 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 5.30 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.119 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay