*** MIDAS logfile (vers. 99NOV) *** Mo, 23 Apr 2001 14:52:16 Page 1 *** MIDAS logfile (vers. 99NOV) *** Mo, 23 Apr 2001 14:52:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010423 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.39 stddev = 2.48 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010423 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1805.7 1804.3 951.9 0.12 0.12 1.90 2 8.23 3556.6 3553.5 1646.7 -0.08 -0.13 2.16 3 16.47 7067.0 7068.9 3246.0 -0.18 -0.14 2.18 4 24.70 10577.5 10579.6 4890.9 -0.17 -0.14 2.16 5 37.04 15844.3 15845.6 7165.3 -0.14 -0.13 2.21 6 49.39 21127.6 21128.0 9781.2 -0.07 -0.07 2.16 7 61.73 26410.6 26411.2 12181.7 -0.01 -0.02 2.17 8 82.31 35233.9 35237.2 16178.4 0.08 0.08 2.18 9 102.89 44074.5 44074.4 20086.6 0.18 0.17 2.19 10 123.46 52932.2 52928.8 24615.3 0.28 0.26 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3919 to 114910 e- Average count rate measured: 427 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.01 e-/ADU RON = 5.39 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.100 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay