*** MIDAS logfile (vers. 99NOV) *** Sa, 14 Apr 2001 14:47:16 Page 1 *** MIDAS logfile (vers. 99NOV) *** Sa, 14 Apr 2001 14:47:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010414 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.31 stddev = 2.48 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010414 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1826.2 1827.1 839.0 0.12 0.15 2.18 2 8.23 3586.8 3583.5 1641.4 -0.07 -0.17 2.19 3 16.47 7111.9 7116.1 3257.3 -0.23 -0.17 2.18 4 24.70 10646.6 10646.5 4920.7 -0.15 -0.15 2.16 5 37.04 15940.0 15942.8 7291.5 -0.14 -0.12 2.19 6 49.39 21251.9 21253.3 9895.9 -0.07 -0.06 2.15 7 61.73 26565.5 26565.0 12395.8 -0.01 0.00 2.14 8 82.31 35436.7 35433.6 16214.7 0.09 0.08 2.19 9 102.89 44327.1 44320.2 20888.5 0.19 0.17 2.12 10 123.46 53228.3 53226.2 24644.3 0.28 0.28 2.16 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3922 to 114288 e- Average count rate measured: 429 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.15 +/- 0.01 e-/ADU RON = 5.31 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.127 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay